BELPYCNO Caractéristiques techniques

Measurement principle Gas displacement method
Sample cell volume 10 cm3, 3.5 cm3, 1 cm3
Measurement accuracy 0.03% of F.S. +0.03% of reading
Repeatability 0.02% of F.S.
Pretreatment Gas purge, flow, vacuum (option)
Pretreatment pressure 0-145kPa (Gauge)
Applicable gas He, N2, inert gas
Measurement pressure 145kPa (Gauge)
Repetition times of measurement Maximum 100 times
Mean number of measurement Maximum 10 times
Measurement temperature range 15-35°C, water bath: 15-50°C (option)
Calibration method Automatic calibration with calibration sphere
Interface Input: RS232C (communication with printer)//Output: RS232C (communication with electric balance)
Dimensions (W x H x D) 270 x 170 x 300 mm
Weight 8 kg
Utility - Gas He, N2, inert gases:150kPa (G)
Utility - Joint 1/8" Swagelok connection
Utility - Power AC100~120V/200-240 V / 200 W, 50 / 60 Hz
CE certified yes