BET Specific Surface Area & Pore Size Analyzer BELSORP MAX G

BET Specific Surface Area & Pore Size Analyzer BELSORP MAX G Technical Data

Measurement principle Volumetric method + AFSM™
Adsorption gas N2, Ar, Kr, CO2, H2, O2, CH4, NH3, NO, CO, butane, and various other (non-)corrosive gases
Gas ports 2 ports (5 ports max.)
Number of measurements//(High Accuracy mode) 1 port in High Accuracy mode
Measurement range (specific surface) 0.01 m2/g and above (N2)//0.0005 m2/g and above (Kr)//(depending on sample density)
Pore size distribution (Diameter) 0.35 - 500 nm
Low pressure isotherm p/p0 = 10^-8//(N2 @77K, Ar @87K)
Pressure transducer 133 kPa (1000 Torr) x 3 units//1.33 kPa (10 Torr) x 1 unit//0.133 kPa (MP) or 0.0133 kPa (LP) x 1 unit
Vacuum gage / pump Turbo molecular pump + rotary pump//Cold cathode gauge (optional)
Sample tube Standard tube, approx. 1.8 cm3 (optional: 5 cm3)
Dewar vessel Volume: 2.6 L//Holding time: 80 h
Pretreatment heater 50 - 450 °C
Water bath -10 - 70 °C
Analysis software BELMaster™ 7 Adsorption isotherm, BET specific surface area I type (ISO9277), BET automatic analysis, Langmuir specific surface area, BJH, DH, CI, INNES method, t-plot, Alpha-s plot
Analysis software BELMaster™ 7 cont. HK, SF, CY method, NLDFT / GCMC, MP method, Dubinin-Astakhov method, Difference adsorption isotherm, Molecular probe, Adsorption rate analysis (opt.)
Dimensions (W x H x D) 320 x 740 x 465 mm
Utility - Gas He, adsorption gas: 0.1MPa (G), purity: more than 99.999%//Joint: 1/8" Swagelok joint
Utility - Power Main unit: AC 100 - 240 V / 850 W, 50 / 60 Hz (including vacuum pump)
ASTM compliance B922, C110, C1069, C1240, C1274, D1993, D3663-20, D3908, D4222, D4365, D4641, D4780, D4842, D5604-96, D6556, D8325, E2864, WK61828, WK71859
USP compliance 268, 846
DIN compliance 66134 (1998-02), 66135-1 (2001-06), 66135-2 (2001-06), 66135-3 (2001-06), 66135-4 (2004-09)
CE certified yes
Recommended Monitor Full HD monitors