Analisador de catalisador BELCAT II

Analisador de catalisador BELCAT II Dados Técnicos

Measurement method Dynamic flow method
Detector Semi-diffusion type 4-element thermal conductivity detector (TCD)
Measurement / pretreatment ports 1
Usable gases He, Ar, N2, O2, H2, CO, CO2, NH3, N2O, NO, etc.
Gas port connections: Total 8
Gas port connections: Carrier gas line 3 out of 8 (non-corrosion resistant lines)
Gas port connections: Mix line 8 out of 8 (including 2 corrosion resistant lines)
Gas port connections: Pretreatment /pulse line 8 out of 8 (including 2 corrosion resistant lines)
Mass flow controller: Carrier gas line F.S. 100 sccm
Mass flow controller: Mix line F.S. 30 sccm
Mass flow controller: Pretreatment / pulse line F.S. 100 sccm
Electric furnace Maximum temperature: 1.200 °C (1.100°C for regular use)//Quick cooling: 30 min (400→50 °C)//Cryo cooling: 10 min (800→30 °C)//CATCryo II: down to -120 °C//Heating rate: 110 °C/min (50 to 500 °C) / 80 °C/min (50 to 1000 °C)
Vapor injection (Option) H2O, CH3OH, C2H5OH, toluene, benzene, etc.
Dimensions (W x H x D), weight 500 x 750 x 500 mm, 80 kg
Requirement: Gas Measurement gas: 0.1 MPa (gauge pressure)//Compressed air: 0.45 to 0.55M Pa (gauge pressure); Joint: 1/8” Swagelok connection
Requirement: Power supply Single-phase, AC110 / 220V
CE certified yes
Vapor (optional): Temperature range Bubbling bottle: Pyrex, 100 cc, 3 to 100 °C, temperature control via Peltier element
CATCryo II (optional): Temperature control method LN2 spray + Heater
CATCryo II (optional): Temperature range Temperature range: -120 ~ 1200°C (1.100°C for regular use)
CATCryo II (optional): LN2 reservoir volume 10 L