High-Precision Gas and Vapor Adsorption Analyzer BELSORP MAX X Series Характеристики
| Measurement principle | Manometric method (volumetric method) + AFSM™ or AFSM™2 |
|---|---|
| Adsorption gas | N2, Ar, CO2, H2, Kr, O2, NH3, NO, CO, CH4, C2H6, C3H8, n-C4H10 , iso- C4H10 and further (non-)corrosive gases |
| Adsorption vapor | H2O/water, C6H6/benzene, C6H5CH3/toluene, CH3OH/methanol, C2H5OH/ethanol, n-C5H12/n-pentane, n-C6H14/n-hexane, c-C6H12/c-hexane, CH2Cl2/dichloromethane, CCl4/carbon tetrachloride and other organic vapors |
| Gas ports | 3 ports (optional: up to 12 ports) |
| Number of measurements | MAX X: max. 4 ports simultaneously//MAX X-HT: max. 4 ports simultaneously//MAXX-HP: max. 3 ports simultaneously |
| Measurement range (specific surface) | 0.01 m2/g and above (N2)//0.0005 m2/g and above (Kr)//(depending on sample density) |
| Measurement range (pore size distribution) | 0.35 - 500 nm |
| Measurement range (gas adsorption isotherm) | MAX X: P/P0 = 10-8~0.997 (N2@77.4K, Ar@87.3K)//MAX X-HT: max. P/P0 = 10-6~0.997 (N2@77.4K, Ar@87.3K)//MAXX-HP: P/P0 = 10-8~0.997 (N2@77.4K, Ar@87.3K) |
| Measurement range (vapor adsorption isotherm) | MAX X: P/P0 = ~0.95 @40°C//MAX X-HT: P/P0 = ~0.95 @80°C//MAXX-HP: P/P0 = ~0.95 @40°C |
| Measurement range (high pressure gas adsorption isotherm) | MAX X: -//MAX X-HT: -//MAXX-HP: 10Pa~900kPa |
| Pressure transducer (1MPa: 7500 Torr) | MAX X: -//MAX X-HT: -//MAXX-HP: 1 unit |
| Pressure transducer (133 kPa: 1000 Torr) | MAX X: 6 units//MAX X-HT: 6 units//MAXX-HP: 5 units |
| Pressure transducer (1.33 kPa: 10 Torr) | MAX X: max. 4 units//MAX X-HT: 4 units//MAXX-HP: 3 units |
| Pressure transducer (0.0133 kPa: 0.1 Torr) | MAX X: max. 3 units//MAX X-HT: -//MAXX-HP: 2 units |
| Thermostatic air oven | MAX X: 50°C//MAX X-HT: 80°C//MAXX-HP: 50°C |
| Vacuum gauge / pump | Cold cathode gage (Optional)/Turbo molecular pump + rotary pump (or diaphragm pump) |
| Measurement temperature (Dewar vessel) | LN2, LAr temperature; Holding time: 80 h |
| Measurement temperature (Pretreatment heater) | 50 - 550°C |
| Measurement temperature (Water bath) | -10 - 80°C (constant temperature circulator) |
| Dimensions (W x H x D) | 360 x 870 x 590 mm |
| Weight (main body) | 50 kg |
| Utility - Gas | He, adsorption gas pressure: 0.1MPa (G)//Joint: 1/8" Swagelok joint//Valve operan gas pressure: 0.5 - 0.6 Mpa (G)// Joint: 1/4" one-touch pipe contact |
| Utility - Power | AC 100 - 240 V/400W (rotary pump not included) |
| CE certified | yes |
| Operating software | Windows 10 operating system or higher, CPU Intel(R) Core i5 or higher//2 GB of memory or more, 5 GB HDD space or more |
| Recommended Monitor | Full HD monitors |