Hochdruck-Gas-Adsorptions-Gerät BELSORP HP

Hochdruck-Gas-Adsorptions-Gerät BELSORP HP Technische Daten

Measuring principle Volumetric gas adsorption method
Adsorptive N2, Ar, O2, CO, H2, CO2, CH4, and other non-corrosive gas (OP:NH3)
Analysis / pretreatment port 1
Pressure transducer Ads isotherm measurements can be achieved by F.S 0.133 MPa and F.S. 13.5 MPa pressure transducers from low to high pressure region.
Measurement pressure range 90% of high pressure transducer F.S.
Temperature range Measurement temp://Water bath + Chiller:-10-70°C (non-freezing liquid for settings below 10°C)//Dewar vessel with LN2 or LAr level controller: 77K, 87K (OP)//Pretreatment temp://50-400°C (OP:50-800°C)
Air circulation bath 15-35°C
Vacuum pump Rotary pump//Ultimate vacuum:<1Pa(7×10-3Torr)
Analysis software BELMaster™ 7 Ad-/desorption isotherm, BET, Langmuir, Horvath-Kawazoe, Saito-Foly, micropore volume by DA, isosteric heat of adsorption, (Clausius-Clapeyron equation), (Option: adsorption rate analysis software)
Dimensions (W x H x D) 600 x 910 x 720 mm
Weight 90 kg
Utility - Gas He: 0.2MPa (G)//Ads gas: Maximum pressure transducer F.S x1.2+0.1MPa (G)//Compressed air: pres. 0.5~ 0.6 MPa (G)
Utility - Joint 1/8" Swagelok connection
Utility - Vent 12mm Ø hose connection (main body, R.P)
Utility - Power AC 100 ~ 120V/200 ~ 240 V, 800 W (excluding vacuum pump) , 50 / 60 Hz
CE certified no
Recommended Monitor Full HD monitors