Wysokociśnieniowy analizator adsorpcji gazów BELSORP HP

Wysokociśnieniowy analizator adsorpcji gazów BELSORP HP Dane techniczne

Measuring principle Volumetric gas adsorption method
Adsorptive N2, Ar, O2, CO, H2, CO2, CH4, and other non-corrosive gas (OP:NH3)
Analysis / pretreatment port 1
Pressure transducer Ads isotherm measurements can be achieved by F.S 0.133 MPa and F.S. 13.5 MPa pressure transducers from low to high pressure region.
Measurement pressure range 90% of high pressure transducer F.S.
Temperature range Measurement temp://Water bath + Chiller:-10-70°C (non-freezing liquid for settings below 10°C)//Dewar vessel with LN2 or LAr level controller: 77K, 87K (OP)//Pretreatment temp://50-400°C (OP:50-800°C)
Air circulation bath 15-35°C
Vacuum pump Rotary pump//Ultimate vacuum:<1Pa(7×10-3Torr)
Analysis software BELMaster™ 7 Ad-/desorption isotherm, BET, Langmuir, Horvath-Kawazoe, Saito-Foly, micropore volume by DA, isosteric heat of adsorption, (Clausius-Clapeyron equation), (Option: adsorption rate analysis software)
Dimensions (W x H x D) 600 x 910 x 720 mm
Weight 90 kg
Utility - Gas He: 0.2MPa (G)//Ads gas: Maximum pressure transducer F.S x1.2+0.1MPa (G)//Compressed air: pres. 0.5~ 0.6 MPa (G)
Utility - Joint 1/8" Swagelok connection
Utility - Vent 12mm Ø hose connection (main body, R.P)
Utility - Power AC 100 ~ 120V/200 ~ 240 V, 800 W (excluding vacuum pump) , 50 / 60 Hz
CE certified no
Recommended Monitor Full HD monitors