Spectromètre de masse quadripolaire
BELMASS II Caractéristiques techniques
| Maximum range |
m/z = 1 ~ 200 |
| Filament material |
Yttria-coated iridium |
| Detector |
Faraday cup / SEM |
| Resolution |
M / -delta-M >= 2M |
| Sniffing section |
Capillary tubes//Standard-specification is made of SUS//Option-specification is made of PEEK |
| Capillary tube dimensions |
OD = 1/16 inch; ID = 0.1 mm; L = 1.5 m |
| Hot hose length |
1 m |
| Maximum heating temperature of hot hose |
200°C (SUS)//120°C (PEEK) |
| Gas introduction method |
Differential exhaust |
| Amount of gas introduced |
0.6 cc / min (at 1 atm) |
| Gas inlet pressure |
Atmospheric pressure |
| Exhaust port fitting |
1/4 inch one-touch joint |
| Communication Interface |
RS-232C (straight) |
| Dimensions (W x H x D) |
280 × 400 × 600 mm (excluding capillary tube) |
| Analog signal input |
DC0 to 10V (10 bit, 1 ch) |
| Weight |
36 kg |
| Power supply: Rated voltage |
Single-phase: 100-120V (Indication 115V)//AC: 200-240V (Indication 230V) |
| Power supply: Power frequency |
50 / 60 Hz |
| Power supply: Power consumption |
600 VA |
| Power supply: Electric shock protection class |
Class I |
| Standard |
CE, UKCA |
| Selected Ion Monitor |
Max. 16 ch |
| Installation environment: Temperature |
10°C to 35°C |
| Installation environment: Humidity |
20%RH to 80%RH (no condensation) |
| Installation environment: Elevation |
2000 m or less |
| Installation environment: Installation category |
Category II |
| Installation environment: Pollution degree |
2 (for indoor use) |